Cipher DGGE Electrophoresis SystemsDGGEK-1001 DGGEK-2001DGGEK-2401 INSTRUCTION MANUAL
10Cipher DGGE Instructions 2/28/142.3.4 Controller Operation 2.3.4.a Reservoir Liquid Level Fill the reservoir with the appropriate buffer. Th
11www.cbsscientic.comTo change from °F to °C, place the Circuit Breaker/Power Switch in the OFF position and then press and hold the P3 Button while
12Cipher DGGE Instructions 2/28/14 2.3.4.h Local Lockout FeatureThis feature enables the user to lock all controls on the controller. While
13www.cbsscientic.com2.3.6 Controller Calibrations and Maintenance 2.3.6.a Calibration Calibration allows the user to match the Controller’s
14Cipher DGGE Instructions 2/28/14DGGEK Unit Set-Up, con’t.2.4 Setting Buffer Temperature Warning: Do NOT turn on Heater/Stirrer until tank has
15www.cbsscientic.comTo mains (wall outlet)To mains (wall outlet)Figure 2-2abcdPower Supply Connection
16Cipher DGGE Instructions 2/28/14SECTION 3Gel Casting Options3.1 Preparation/Cleaning of Glass Plate for Gel CastingHand wash both plates with a hig
17www.cbsscientic.com3. Place the gasketed plate on the lab bench with the tubing side up, and extend the bottom of the plate over the edge of the
18Cipher DGGE Instructions 2/28/143.2 Gel Casting Techniques- continuedB. Vertical gradient gel casting using GM-40 gradient maker and gravity ow.1.
19www.cbsscientic.com3.2 Gel Casting Techniques- continuedC. Vertical gradient gel casting using gradient maker (GM-40) and a mini-pump. 1. Alter
2Cipher DGGE Instructions 2/28/14TABLE OF CONTENTS Important User Information . . . . .
20Cipher DGGE Instructions 2/28/143.2 Gel Casting Techniques- continued D. Vertical gradient gel casting using GM-40 gradient maker, Mini-Pump and
21www.cbsscientic.com3.3 Vertical Gradient Formation To determine the range of gradient appropriate for your fragment analysis, please read the en
22Cipher DGGE Instructions 2/28/14FIGURE 1FIGURE 21.0mm AND THICKER.75mm AND THINNERPERPENDICULAR GEL CASTINGTO GRADIENT MAKER20 GAUGE NEEDLESEAL WITH
23www.cbsscientic.com3.5 Perpendicular Gradient FormationTo determine the range of perpendicular gradient appropriate for your fragment analysis, p
24Cipher DGGE Instructions 2/28/14SECTION 4 Gel Running Preparations4.1 Preparation of the Cassettes1. After polymerization rinse gel plate assembl
25www.cbsscientic.com4.2 Buffer Cycling Connections1. Fill upper reservoir(s) by attaching recirculating tubing to barb tting on each upper reser
26Cipher DGGE Instructions 2/28/144.2 Buffer Cycling Connections-continued.Buffer Cycling Connections – Dual Cassettes SECTION 5Running C
27www.cbsscientic.com5.2 Removing the Gels3. After the run is completed, turn off power supply, disconnect recirculating tubes and power leads, remo
28Cipher DGGE Instructions 2/28/14SECTION 6Recommended Buffers and Reagents Warning: Do not mix buffer in the DGGE tank! 6.1
29www.cbsscientic.comSECTION 7Maximum Well/Comb Volumes Vertical Comb Options, deeper wells, overall length = 13.1cmCat. # Tooth Depth(mm)# of T
3www.cbsscientic.comIMPORTANT USER INFORMATIONThis Instruction Manual will explain how to use this product safely and effectively. Please read and c
30Cipher DGGE Instructions 2/28/14SECTION 8 References1. S.G. Fischer and L.S. Lerman, (1983) PNAS 80:1579.2. R.M. Myers, T. Maniatis, L.S. Lerman,
31www.cbsscientic.comSECTION 9Troubleshooting 9.1 Unit Will Not Operate (no heat, cooling, or pumping) • Check that the power cord is plugged in to
32Cipher DGGE Instructions 2/28/14SECTION 10Maintenance of Equipment10.1 Care and HandlingThe plastic components of the DGGE Systems are fabr
33www.cbsscientic.comSECTION 11Ordering Information CAT. # DESCRIPTIONCIPHER DGGE ELECTROPHORESIS SYSTEMSDGGEK-1001Cipher DGGE Electrophoresis Syst
34Cipher DGGE Instructions 2/28/14
35www.cbsscientic.com 13R, 712, 7306, 7312, 812, 8006, 8012, 8002, 8112, 8102, 8106, 8202, 8212, 8206, 8306, 9006, 9012, 9002, 912, 9112, 9102, 91
CONTACT INFORMATIONOnline Ordering www.cbsscientic.comTelephone Local and International: 858-755-4959 Toll Free: 800-243-4959Sales E-mail Address
4Cipher DGGE Instructions 2/28/14FRANÇAIS INFORMATION IMPORTANTE À L’USAGE DES UTILISATEURSLe présent manuel d’utilisation explique la manière de se
5www.cbsscientic.comSECTION 1General Information 1.1 IntroductionDenaturing Gradient Gel Electrophoresis (DGGE) is a powerful genetic analysis
6Cipher DGGE Instructions 2/28/14 1.3 SafetyPower to the DGGE systems is to be supplied by an external DC voltage power supply that must be grou
7www.cbsscientic.comFigure 2-1 HeaterStirrerBuffer SiphonPump AssemblyVaporShieldThermometerDGGETankSingleCassetesComb (2
8Cipher DGGE Instructions 2/28/14 Front View Rear View 1. Dis
9www.cbsscientic.com 1. Temperature Sensor 2. Pump Shaft and Impeller 3. Pump Outlet Nozzle1 4. Over-Temperature S